Cryogenic Manifolds, Liquid Withdrawal - Liquid Dispense

Contact us

CFAM-PX Series

Switchover Process Actuated by Pressure

How it works The controller monitors the pressure of each liquid cylinder. Once the liquid cylinder in service is low in pressure, it will switch over to the other liquid cylinder.  
Benefits Extremely fast reaction time, delivers cryogenic fluid at the minimum pressure required by the end user, up to 500 psig.
Draw back Not suitable to supply liquid nitrogen to cryogenic freezers – does not monitor how much liquid is left in the cylinder or the temperature of the fluid delivered.
Typical application

Perfect for any applications requiring high flow of gas such as hyperbaric chambers or laser cutting machines.

Specification Sheet
Instruction Manual
CFAM-WX Series

Switchover Process Actuated by Weight

How it works The controller monitors of weight of each liquid cylinder. Once the weight of the liquid cylinder in service falls below a target weight, it will switch over to the other liquid cylinder.  
Benefits Fast reaction time, can work with any type of cylinder size or style, designed to handle very high or very low pressure, and allows the use of pretty much all cryogenic liquid in cylinder.
Draw back Because of the two scales, it is the most expensive system of the three. It also takes the most floor space. For some operators, pushing the heavy liquid cylinders on the scales can be challenging.
Typical application

No limit. The limitation is the capacity of the liquid cylinders to perform the desired task.

Specification Sheet
Instruction Manual
CFAM-TX Series

Switchover Process Actuated by Temperature and Pressure

How it works The switchover from a liquid cylinder in service to a stand-by liquid cylinder will occur once both liquid cylinder pressure and fluid temperature conditions are not met over a certain period of time.
Benefits Because the system does not use scales, it has some of the same benefits of the CFAM-WX (minus the reaction time) without the huge price tag.
Draw back Reaction time – two minutes can pass before the controller determines that the temperature and pressure conditions are no longer within “acceptable” parameters.
Typical application

Geared toward cryogenic freezers provided that either the freezers or the CFAM-TX are equipped with the hot-gas-by-pass option.

Specification Sheet
Instruction Manual

LCFS Series

Cryogenic Liquid Cylinder Filling Station

Purpose On-site cryogenic liquid cylinder filling station
How It Works Draws cryogenic liquid from bulk tank to fill cryogenic liquid cylinders
Fluid Service Liquid Argon, Liquid Nitrogen, Liquid Oxygen
Maximum Working Pressure 150 Psi
Specification Sheet 
Instruction Manual