When it comes to medical air, there is no room for error or compromise. The quality and performance of your medical air system is of crucial importance, as is correct installation. BeaconMedaes’ TMA Medical Air System assembled as a final medical product for proper and fast commissioning, as well as optimal performance.
100% redundancy
The TMA comes equipped with duplex medical air compressors, medical air dryers, coolers, controllers and other components to protect from single fault failure.
Fully connected and factory-tested control system
The TMA control system is integrated, wired, connected and tested at the factory. This guarantees correct wiring and control set-up.
Safe, easy service
Compartments are isolated mechanically, pneumatically and electrically on each side to enable safe service interventions on one part while the other part keeps running.
The TMA comes as a duplex unit in one canopy. That means it can be used for surgical (HTM), instrument (NFPA) and non-medical air applications. Only the receiver must be piped on-site to complete the installation. A back-up manifold serves as a secondary supply source.
HTM02-01 specifies that the continuity of a medical air system should be provided by three sources: three compressors and one back-up manifold. The GA VSDs will serve as the primary supply source, the TMA’s two integrated compressors will be the secondary source and the back-up manifold will be the third. Using one external VSD compressor gives you maximum energy savings with a lower capital investment.